In the last lecture, I touched on

0:03

the principles of contrast injection.

0:06

There were some very busy slides,

0:07

so I want to introduce those.

0:12

give you the fuller view of what's going on.

0:16

So first, what happens is that a field of view

0:22

is prescribed, a pre-monitoring slice is chosen, a

0:26

region of interest is placed on it, and then the

0:30

monitoring slice is based on that region of interest.

0:35

Then what happens is that the tech goes

0:38

inside the room, checks and makes sure that

0:40

everything is hooked up properly, and comes

0:43

out of the room and then presses two buttons.

0:45

One alerts that the CT scan is about to start,

0:49

and the other actually starts the contrast.

0:52

So let's say T equals zero, contrast starts.

0:57

So now your contrast is in the veins and it's going

0:59

through the right heart and coming into the left heart.

1:03

At around 10 seconds, you start the monitoring slice.

1:10

In 10 seconds, maybe a little early.

1:13

There may sometimes be situations

1:14

where the circulation is hyperdynamic.

1:17

So it starts looking at the descending aorta and

1:21

taking these low-dose scans with a region of interest.

1:25

And the monitoring of the bolus starts.

1:27

Around 22 seconds, the attenuation in the

1:30

descending aorta reaches our preset, which is 130.

1:34

It can be lower, it can be higher.

1:37

And then, after a trigger delay, let's say 6 seconds,

1:42

maybe 10 seconds or 12 seconds, the acquisition starts.

1:47

And then the breath hold instruction is built

1:50

in so that it's factored in within that delay.

1:53

So let's say at about 28 seconds the

1:55

acquisition starts and at about 34 seconds.

1:57

So you can see that in a scan that is only

2:01

6 seconds long, the whole process, the

2:04

whole orchestra has to be about 34 seconds.

2:09

And most of that time is not scanning; most of

2:12

the time is just the injection getting there.

2:15

That's important to understand, um, because our

2:19

tendency is to think of this as being real time.

2:21

It's not; it's a snapshot of what's going

2:25

on once the injection has been made.

2:28

So what can be the issues with this?

2:31

Well, the first issue is if you

2:32

have a lot of atherosclerosis

2:35

in the descending aorta, or even the ascending

2:38

aorta, which would mean that there's, the

2:41

lumen is not really a lumen; it's black.

2:44

Similarly, if you have a dissection, and accidentally place

2:48

the ROI into the false lumen, that can be a problem.

2:53

You can have streak artifact from metal that can

2:55

falsely elevate the attenuation value, which can

3:00

falsely start the study, or you could have very

3:02

low cardiac output that it takes such a long time

3:05

to reach the descending aorta that the scanner has

3:08

this automatic system where it just scans and then

3:11

you get a suboptimal study as a result of that.

3:15

So how.

3:18

Would you vary?

3:19

I mean, there are a few things in

3:19

your control with this bolus tracking.

3:23

If you want to scan later, so you have

3:25

these presets that are there in the scanner

3:27

and they're based on crude experience,

3:32

but sometimes there's a variation theme.

3:34

You might want to scan earlier,

3:35

you might want to scan later.

3:37

So if you want to scan later,

3:38

there are two things you can do.

3:39

You can either increase the threshold hands

3:41

free unit, let's say from 130 to 180, and I

3:44

honestly, I don't think that does that much.

3:47

If you look at the threshold, it kind

3:48

of goes 90, 130, and it really jumps up.

3:52

The better thing to do is to increase the

3:54

trigger delay; whatever that trigger delay was.

3:57

If you're worried about starting too

3:59

soon, you increase that trigger delay.

4:01

And remember, trigger delay is

4:02

not the same as monitoring delay.

4:04

And also remember, trigger delay

4:05

is not always called trigger delay.

4:06

So just understand what those terms are.

4:09

If you want the acquisition to start

4:11

sooner, you have a couple of options.

4:13

One is to reduce the hands-on unit.

4:15

And again, I think that that is going to be problematic

4:18

because if you reduce it too much, you can be in the

4:21

noise territory where it'll just spuriously start.

4:25

Or you can reduce the trigger delay.

4:27

And that, again, is something that you can

4:29

only do to an extent because there's a certain

4:33

inbuilt minimum that you need to rev the scan up.

4:38

And again, monitoring delay

4:39

is not going to do very much.

4:40

I mean, of course, you can not monitor for

4:43

another 10 seconds, but you might miss the peak.

4:45

It doesn't determine when to start.

4:47

That determines when you're looking at it.