When there is ECG synchronization, there can be
0:03
problems with the synchronization either because
0:06
patients has arrhythmia, ectopic beats, or the scanner
0:11
fails to recognize the, um, uh, certain portions
0:15
of the cardiac cycle, notably the, uh, R wave.
0:18
So these are pretty common and sometimes they lead
0:23
to dramatic action like getting the patient back.
0:26
Often all they need is just a little bit of editing.
0:30
So here's an example of a patient who
0:32
came in for a pre-op evaluation for a
0:37
TAVR, the endovascular aortic valve.
0:41
And in that we get multi-phase
0:43
images of the aortic root and aorta.
0:50
And you can see here quite clearly that there
0:53
is a lot of artifacts, motion artifacts.
0:57
And the thing to do then is to go to the scanner and
1:02
look at the raw data and look at the original ECG.
1:06
The original ECG you can get by looking at these tabs.
1:13
Of course, it depends on which type of scanner you have.
1:15
And you can get something like this,
1:17
a rhythm strip, a single rhythm strip.
1:19
And you can see here what the problem is.
1:21
We've got a nice little block here,
1:22
a nice little block over here, a nice
1:24
little block over here, a nice block here.
1:26
And then in between we've got PVC.
1:30
That's confused.
1:32
The, um,
1:34
scanner, and there is a compensatory pause.
1:37
That's the reason why you have
1:39
that motion out of bounds.
1:41
And in order to deal with this,
1:42
you have to really set yourself.
1:44
Okay, what should the patients
1:46
rhythm have been, had they not been
1:48
that PVC and then insert a new kind of
1:51
heartbeat, an artificial heartbeat.
1:54
And when you do that, you can get
1:57
dramatic changes in the image quality.
2:03
This and this may be difficult to compare,
2:05
but you can see that this and this,
2:09
dramatic changes in the image quality.
2:15
So here's another example.
2:17
Again, this is a patient who came in for a pre-
2:21
op TAVR, and this is the multi-phase image.
2:26
No doubt, there is a lot of ghosting.
2:32
And this is quite dramatic, and this is a
2:35
problem because if you have ghosting you can
2:40
overestimate the annular area and then end up
2:44
implanting a valve that's too big for the patient.
2:49
which can have its own consequences.
2:51
So, again, you can see what the issue is.
2:54
Looking at the ECG strip, there's just
2:56
this one area that just got, um, missed.
3:00
And, when we correct that, we can see very nice,
3:05
crisp images which are a result of the ECG correction.
3:11
This is also an interesting one,
3:14
which, see this band over here?
3:17
This band is because of an abnormal beat.
3:20
And, let's see if you can spot that abnormal beat.
3:23
Actually, you don't have to because what we've
3:25
done is we've taken that abnormal beat out.
3:28
See this line over here?
3:30
That's taking it out, deleting it or disabling it.
3:34
You can do either one.
3:36
And you can see what has happened as a result of that.
3:39
It's a dramatic change in the image quality.
3:42
And even though we took it out, we actually made this
3:47
heart rate more irregular, but it's better to have.
3:51
Irregular and without the PVC because
3:54
PVC doesn't really follow the same course.
3:57
It's just something that confuses the system.
4:00
So small changes resulting in dramatic differences.
4:04
And here's an example again of ECG editing.
4:09
We've got motion in the RCA that looks like it
4:12
may be incorrectable, but you can see that you
4:15
go back to the ECG and you can see a couple
4:17
of problems where the arrow is pointing.
4:21
of these abnormal beats, and what we've done
4:27
there is just simply disabled the sync, and that's
4:31
led to dramatic improvement in imaging quality.